3/5/60 C. McIlwain -- L. Frank X-ray machine check: At I ~= 5.7 milliamperes, the lights on the X-ray control panel read 17 kv low for a "preset" meter reading of 6.0 ! This is the situation which we have habitually used in X-ray calibrations during the past several years. All of our old kv readings